CAUTION: DISCONNECT ALL ELECTRICAL CONNECTIONS BEFORE
PROCEEDING WITH DISASSEMBLY AND ASSEMBLY
1. Disconnect the 1/50HP motor from the existing magnetic drive assembly by unscrewing the
upper cover (2982HC) from the end cap (2717HC2) as shown in Picture 1.
2. Loosen set screw in drive adapter (2938HC) and remove the drive adapter from the motor
shaft. Set the motor aside.
3. Remove the stirrer shaft from existing magnetic drive assembly
4. Remove the magnetic drive assembly from the reactor head using a 264AC4 spanner wrench.
5. Some external fittings may need to be removed for easier removal of the magnetic drive
assembly.
444M
ASSEMBLY INSTRUCTIONS FOR
5500 RETRO-FIT
2982HC
Upper Cover
2938HC
Drive Adapter
2717HC2
End Cap
Existing
Motor
Existing Magnetic
Drive Assembly
Picture 1
Seitenansicht 0
1 2 3 4 5 6 7 8

Inhaltsverzeichnis

Seite 1

1 CAUTION: DISCONNECT ALL ELECTRICAL CONNECTIONS BEFORE PROCEEDING WITH DISASSEMBLY AND ASSEMBLY 1. Disconnect the 1/50HP motor from the

Seite 2

2 ASSEMBLING THE 5500 RETRO-FIT 1. Unscrew the (3) 8-32 x 3/8 BHSCS from the Top Cap (3078HC) to separate the Top Cap from the (3) support rods (

Seite 3 - Rotate bottom cap

3 5. With ears of bottom cap clear of fittings continue to screw in the retro-fit assembly into the head of the reactor and tighten with the

Seite 4 - Support Rod

4 8. Align the holes in the Top Cap (3078HC) with the (3) support rods and install the (3) 8-32 X 3/8 BHSCS with the 3/32” socket screw key pro

Seite 5

5 9. Make sure the bottom cap is seated flat on the ring of the magnetic drive assembly body as shown in picture 4 and tighten the (2) 8-32 x 1/4

Seite 6 - 3083HC Coupling Half

6 11. Screw existing motor onto the top cap of retro-fit assembly, aligning the coupling half on the motor with the b

Seite 7

7 12. Check set screw of the coupling on drive adapter and tighten if necessary. With the correct installation of all parts there should be a

Seite 8 - Revision 1-5-04

8 14. Upon completion of assembling, place reactor in base/stand and connect all electrical connections to tes